Ekipmanlar

  • DURRIDGE RAD7

    Principle of Operation

    Electrostatic collection of alpha-emitters with spectral analysis

    Passivated Ion-implanted Planar Silicon detector

    SNIFF mode counts polonium-218 decays

    NORMAL mode counts both polonium 218 and polonium 214 decays

    Built-In Air Pump

    Nominal 1 liter/minute flow rate

    Inlet and outlet Luer connectors

    Connectivity

    RS-232 port up to 19,200 baud rate

    USB adaptor is included with every RAD7

    Measurement Accuracy

    +/-5% absolute accuracy, 0% - 100% RH

    Nominal Sensitivity

    SNIFF mode, 0.25 cpm/(pCi/L), 0.0067 cpm/(Bq/m3) NORMAL mode, 0.5 cpm/(pCi/L), 0.013 cpm/(Bq/m3)

    Radon Concentration

    Range

    0.1 - 20,000 pCi/L (4.0 - 750,000 Bq/m3)

    Intrinsic Background

    0.005 pCi/L (0.2 Bq/m3) or less, for the life of the instrument

    Recovery Time

    Residual activity in Sniff mode drops by factor of 1,000 in 30 minutes

    Operating Ranges

    Temperature: 32° - 113°F (0° - 45° C) Humidity: 0% - 100%, non-condensing

    Cycle Range

    User controllable number of cycles, from 1 to 99 to unlimited, per run

    User controllable cycle time, from 2 minutes to 24 hours