Principle of Operation
|
Electrostatic collection of alpha-emitters with spectral analysis
Passivated Ion-implanted Planar Silicon detector
SNIFF mode counts polonium-218 decays
NORMAL mode counts both polonium 218 and polonium 214 decays
|
Built-In Air Pump
|
Nominal 1 liter/minute flow rate
Inlet and outlet Luer connectors
|
Connectivity
|
RS-232 port up to 19,200 baud rate
USB adaptor is included with every RAD7
|
Measurement Accuracy
|
+/-5% absolute accuracy, 0% - 100% RH
|
Nominal Sensitivity
|
SNIFF mode, 0.25 cpm/(pCi/L), 0.0067 cpm/(Bq/m3) NORMAL mode, 0.5 cpm/(pCi/L), 0.013 cpm/(Bq/m3)
|
Radon Concentration
Range
|
0.1 - 20,000 pCi/L (4.0 - 750,000 Bq/m3)
|
Intrinsic Background
|
0.005 pCi/L (0.2 Bq/m3) or less, for the life of the instrument
|
Recovery Time
|
Residual activity in Sniff mode drops by factor of 1,000 in 30 minutes
|
Operating Ranges
|
Temperature: 32° - 113°F (0° - 45° C) Humidity: 0% - 100%, non-condensing
|
Cycle Range
|
User controllable number of cycles, from 1 to 99 to unlimited, per run
User controllable cycle time, from 2 minutes to 24 hours
|